Data
98,375 bulk items are available, covering 93,242 participants.
Defined-instances run from 2 to 3, labelled using Instancing 
2.
The values held within this field are coded links to files held within a bulk repository, not the actual data itself.
 
2 Instances
Instance 2 : Imaging visit (2014+)
93,177 participants, 93,177 items
Data not meaningful for statistical summary
Instance 3 : First repeat imaging visit (2019+)
5,198 participants, 5,198 items
Data not meaningful for statistical summary
 
Notes
The DICOM images were acquired using the following parameters: 
 Resolution: 2x2x2 mm 
Field-of-view: 104x104x72 matrix 
Duration: 7 minutes (including 36 seconds phase-encoding reversed data) 
10x b=0, 50x b=1000, 50x b=2000 
SE-EPI with x3 multislice acceleration, no iPAT, fat saturation 
 For the two diffusion-weighted shells, 50 distinct diffusion-encoding directions were acquired (and all 100 directions are distinct). The diffusion prepraration is a standard ("monopolar") Stejskal-Tanner pulse sequence. This enables higher SNR due to a shorter echo time (TE=92ms) than than a twice-refocused ("bipolar") sequence. This improvement comes at the expense of stronger eddy current distortions, which are removed in the processed data available in NIFTI format.
 
2 Related Data-Fields
| Field ID | Description | Relationship | 
| 25780 | Acquisition protocol phase. | Field 25780 indicates the protocol used to acquire Current Field | 
| 20250 | Multiband diffusion brain images - NIFTI | Field 20250 is derived from Current Field |